The Effect of Mismatch of Total Knee Replacement Components with Knee Joint : A Finite Element Analysis
Abstract
It has been noticed that the need for total knee replacement surgery is increasing for Asian region. A total knee replacement is a permanent surgical solution for a patient having debilitating pain in knee joint suffering from arthritis. In this surgery, knee joint is replaced with components made up of bio-compatible materials after which the patient can resume the normal day to day activities.
Western population has bigger build compared to Asian population. Most of the total knee replacement prosthesis are designed for western population. When these total knee prosthesis are used for Asian population, they cause a mismatch leading to various clinical complications such as reduced range of motion and pain. The studies have been limited to clinical complications caused by the mismatch. To address this limitation, current study is aimed to find the mechanical implications such as stress distribution, maximum stresses, maximum displacements etc., caused by mismatch of total knee replacement components with knee. A surgeon selects total knee components for a patient based on some critical dimensions of femur and tibia bone of knee. In addition, a method to accurately calculate these dimensions of the femur and tibia bone of a real knee was developed in the current study. This method calculated the points of curvature greater than a threshold (decided based on the radius of the curvature) found out using the formula of curvature. Further, the highest point was calculated based on maximum height from a line drawn between initial and final point within the captured points, also the extreme points were calculated based on the sign change in slope of points within the captured points, giving multiple points on the boundary of bones extracted in an MRI image of a patient. The distance between two selected farthest points, out of these points, in specific direction was the basis for selection of the TKR components.
Total knee replacement components were modeled in Geomatics Studio 12 software, bones were modeled in Rhinoceros 5 software, assembly of bones and total knee replacements components was done in Solid works 2013 software, the finite element model of the assembly was developed in Hyper mesh 11 software and, the stress analysis and post processing was done in ABAQUS 6.13 software. A static, implicit non linear analysis was performed. Simulations were performed for two conditions: at standing (0o of flexion) and at hyper-flexed (120o of flexion). In order to figure out if there were any mechanical implications of mismatch, the full model of assembly consisting of femur, tibia and fibula bones assembled with total knee replacement components, and the reduced model consisting of only total knee replacement components were simulated separately, results of which have been discussed in the current thesis. In this work, the effect of change of length of ligaments at 120o of flexion in detail was also studied. This study brought out various outcomes of contact mechanics and kinematics between the components of total knee replacement prosthesis.
Collections
Related items
Showing items related by title, author, creator and subject.
-
Cache Coherence State Based Replacement Policies
Agarwal, Tanuj Kumar (2018-07-14)Cache replacement policies can play a pivotal role in the overall performance of a system by preserving data locality and thus limiting the o -chip accesses. In a shared memory system, a cache coherence protocol is necessary ... -
Structure Determination of Proteins of Unknown Origin by a Marathon MR Protocol and Investigations on Parameters Important for Molecular Replacement Structure Solution
Hatti, Kaushik S (2018-02-20)Occasionally, crystallisation of proteins works in mysterious ways! One might obtain crystals of a protein of unknown identity in place of the protein for which crystallisation experiments were performed. If the investigator ... -
Structural Studies On Enzymes From Salmonella Typhimurium Involved In Propionate Metabolism: Biodegradative Threonine Deaminase, Propionate Kinase And 2-Methylisocitrate Lyase
Simanshu, Dhirendra Kumar (2007-11-23)I formally joined Prof. M. R. N. Murthy’s laboratory at the Molecular Biophysics Unit, Indian institute of Science, on 1st August 2001. During that time, the interest in the laboratory was mainly focused on structural ...